Specifications and functions |
Dimension | 360 x 145 x 123mm |
Syringe size | 5, 10, 20, 30, 50/60ml |
Infusion mode | Rate mode |
Flow rate range | 5ml syringe: (0.1~150)ml/h; 10ml syringe: (0.1~300)ml/h; 20ml syringe: (0.1~600)ml/h 30ml syringe: (0.1~900)ml/h; 50ml/60ml syringe: (0.1~1500)ml/h |
Accumulated volume | 0.1〜9999 ml |
VTBI | 0.1~999.9 ml, increment is 0.1 ml; 1000-9999ml, increment is 1ml |
Increment | 0.1~999.9 ml/h, increment is 0.1 ml/h; 1000~1500 ml/h, increment is 1 ml/h |
Accuracy | ≤ ±2% |
KVO rate | 0.5 ml/h |
Bolus rate | 0.10-1500ml/h |
Purge rate | 150-800ml/h |
Occlusion | Low 300±125mmHg(40kPa);Medium 525±125mmHg(70kPa); High 900±180 mmHg (120 kPa) |
Screen | 2.5'' LCD monochrome screen, 132 x 32 pixels |
Tit-ration | Yes, increment is 0.1ml/h |
Keypad lock | Yes |
Pause | Yes |
Automatic recognition of syringe size | Yes |
Self-test system | Yes |
Weight | <1.95kg (not including clamp) |
Handle | Embedded handle |
Mounting type | Standard: Vertical Pole |
Alarm Volume | 1-8 levels selectable |
Preset syringe brand | Include B-D Luer-Lok Tip,B.Braun OPS,Terumo SS; Local brand configuration supported |
Language | 7, Chinese, English, Spanish, French, Russian, Turkish, Portuguese |
Alarm |
| VTBI done, VTBI near done, KVO finish, syringe empty, syringe near empty, syringe not installed, syringe disengaged, occlusion, system error, system abnormal, battery low, battery empty, reminder, AC power disconnection, standby time expired |
Safety Requirement |
AC power | 100-240V (50/60Hz, 0.4-0.14A) |
DC power | 10-16V (2.0-1.25A) |
Battery type | Lithium |
Battery operation time | >6 hours@5ml/h |
Charging time | <5 hours@100% |
Classification | CF, Class I |
IP Grade | IP 34 |
Temperature | 5~40°C for operating, -20~60°C for storage |
Humidity | 15~95% for operating, 10~95% for storage |
Air pressure | 57~106kPa for operating, 50~106kPa for storage |
Interface |
Data interface | Supported, RS232 |
Nurse Call | Supported |
DC input | Supported |