Feature:
1.CPAP ventilation mode keeps children spontaneous breath and reduces children
breathing work;
2. Provide CPAP mode of constant flow and limited pressure for newborn baby;
3. Offer manual ventilation mode of pressure limit, doctors can operate by the buttons
replacing manual bag ventilation to save treatment time;
4. Delivery air/ O2 mixture with properly adjusted O2 concentration and continuous
positive pressure to premature baby, newborn baby and child, which will highly improve
the safety and efficacy of oxygen therapy;
5. Equipped with built-in oxygen/air blender ,ensuring O2 concentration adjustment
to be stable;
6. High quality oxygen sensor can monitor real-time O2 concentration , alarm over limitation, ensure safety of breathing O2;
7. Offer manual ventilation mode of pressure limit, doctors can operate by the buttons
replacing manual bag ventilation to save treatment time.
8. Reliable CPAP control valve and advanced pressure monitoring system improve
CPAP adjustment precision.
Specification:
Ventilation mode | |
CPAP, NIPPV, HFNC, MANUAL, NCPAP | |
Parameter | |
Oxygen concentration | Adjusting range: 21% ~ 100% Monitoring range: 15% ~ 100% |
CPAP | Adjusting range: 0 cmH2O ~ 15 cmH2O Monitoring range: 0 cmH2O ~ 60 cmH2O |
Continuous flow | Adjusting range: 2 L/min ~ 20 L/min Monitoring range: 0 L/min ~ 20 L/min |
Pressure monitoring | 0 cmH2O ~ 60 cmH2O |
Inhalation pressure | 3 cmH2O ~ 15 cmH2O |
Exhalation pressure | 1 cmH2O ~ 13 cmH2O |
Breath Frequency | Adjusting range: 1 /min ~ 100 /min Monitoring range: 0 ~ 100 /min |
Inhalation time | Adjusting range: 0.1 s ~ 15 s Monitoring range: 0 s - 15 s |
Manual ventilation | Output flow: 20 L/min ± 5 L/min |
Air compressor | Continuous output flow: ≥ 25 L/min Peak output flow: ≥ 120 L/min |
Oscillogram | P-T (pressure – time) Flow of oxygen, air and mixed gas Airway pressure oscillogram |
Alarm and protection | |
The AC power failure alarm | Power failure or no connection |
Internal battery backup low voltage alarm | < 11.3 ± 0.3 V |
High oxygen concentration alarm | 19% ~ 100% |
Low oxygen concentration alarm | 15% ~ 99% |
High Airway pressure alarm | 1.5 kPa ~ 6k Pa |
Low air input pressure alarm | < 200 kPa |
Low oxygen input pressure alarm | < 200 kPa |
Air and oxygen blender alarm | When any gas source lack When differential pressure > 0.1m Pa |
Compression pump alarm Working condition | Excess temperature |
Gas source | O2, Air |
Pressure | 280 kPa ~ 600 kPa |
Voltage | 100 ~ 240 V |
Power frequency | 50 Hz ± 1 Hz |
Wooden case packing size | |
Main components | L 683 * W 687 * H 1140 mm |
G.W | |
N.W. | |
Air compressor | L 505* W 566 * H 615 mm |
G.W. | 43 KG |
N.W. | 22 KG |