Feature:
3/5 lead ECG,suppot 7 ECG waveforms
Excellent ECG Anti-ESU, anti- defibrillator performance
Unique NIBP algorithm keep accuracy in all kinds of patient conditions
SPO2 have DSP algorithm and excellent performance when movement;
720 hours trend data and 2 hours wave data
Unique human voice alarm
USB WIFI module can easy connect to CMS
7 screen layouts: Standard, BigFont, Trend, OxyCRG, ECG Full-lead, OtherBed;
Dynamic screen setting, every paramenters and waveforms can set open or closed and set color
USB data output with software to view data on computer
Gas module can auto detect 5 anethesia gas type: HAL, ENF, ISO, SEV, DEV
Suport 16 languages include: English, French, German, Spanish, Portuguese, Russian and etc
USB data output with software to view data on computer
Moldular monitor has 4 slots inlucde: IBP, CO2, GAS and CSM/IoC
Specification:
Quality Standards and Classification
CE.ISo13485
SFDA: CLASS II B
Anti-electroshock degree
Class i equipment
(internal power supply)
TEMP/SPO 2/NIBP BF
EGG/RESP:CF
Application range
Adult/pediatric/neonatal/medicine/surgery/operating room/ICU/CCU
Display
8"real color TFT screen
Environment
Operating environment :
Temperature: 0 ~ 40 ℃
Humidity: ≤85%
Altitude: -500 ~ 4600m
Transport and Storage environment :
Temperature: -20~ 60 ℃
Humidity: <93%
Altitude: -500 ~ 13100m
Power Requirements
AC: 100 ~ 240V, 50Hz/60Hz
DC: Built-in rechargeable battery
Battery: 4000mA 11.1V lithium battery
2h operating after full charge (one piece)
5min operating after low battery alarm
Dimension and Weight
Equipment:
214 mm x 157 mm x238 mm 2 kg
Package:
380 mm x350 mm x 300 mm 4.3 kg
Date Storage
Trend diagram/table: 7X24h
NIBP review: 400 events
Wave review: 100min
Alarm review: 100 alarm events
Support drug concentration titration analysis
ST Segment
ST Segment Range: -2.0mV ~ +2.0mV
Accuracy:0.02mv
RESP
Method: RA-LL impedance
RESP Impedance range: 0.3~3Ω
Base Impedance range: 200 Ω ~ 4000 Ω
Bandwidth: 0.1 ~ 2.5Hz
RESP rate: Adult: 0~ 120BrPM
Neonatal/Pediatric : 0 ~ 150BrPM
Resolution : 1BrPM
Precision: ±2 BrPM
Asphyxia Alarm: 10~40 seconds
ECG
5 Leads: RA, LA, LL, RL, V
Lead mode: I, II, III, AVR, aVL , aVF, V
Increase:x250、x500,x1000,x2000
Sensitivity: >200 uV( Peak-to-peak value)
Input impedance: >5(megohm)
Bandwidth: Surgery 1-20 Hz
Monitor 0.5-40 Hz
Diagnostic 0.05-130 HZ
CMRR:≥100db
Polarization Voltage: ±300mv
Baseline Recovering Time: After defibrillation <3 seconds
Signal Range: 8 mV(peak-to-peak value)
Calibrating Signal: 1mV, precision ±5%
NIBP
Method: Pulse wave oscillometry
Work mode: Manual/ Auto/ STAT
Measure interval of auto mode:
1,2,3,4,5,10,15,30,60,90,120,180,240,480 minute(s)
Measuring Time of STAT Mode: 5 minutes
PR range: 40 ~ 240bpm
Measure & alarm range: Adult
SYS 40~270mmHg
DIA 10 ~ 215mmHg
MEAN 20~235mmHg
Pediatric
SYS 40 ~ 200mmHg
DIA 10~ 150mmHg
MEAN 20~ 165mmHg
Neonatal
SYS 40 ~ 135mmHg
DIA 10 ~ 100mmHg
MEAN 20~110mmHg
Static pressure range: 0~ 300mmHg
Precision: ±3mmHg
Pressure precision:
Max. average error: ±5mmHg
Max. standard deviation: ±8mmHg
Overvoltage protection:
Adult: 300mmHg
Pediatric: 240mmHg
Neonatal: 150mmHg
Spo2
Range: 0-100%
Resolution: 1%
Precision: 70%-100%: ±2 DIGIT
0%-69%: ±no definition given
Pulse Rate
Range: 20-300bpm
Resolution: 1bpm
Precision: ±3bpm
TEMP
Channel: 2
Measure & alarm range: 0-50℃
Resolution: 0.1℃
Precision (no sensor): +0.1℃
Standard accessories
NIBP cuff& tube
ECG cable & electrodes
Spo2 sensor
TEMP probe
Lithium-ion battery
Power cable
Operator’s manual
Optional accessories
Co2 module
IBP module
Trolley bracket
Hanging bracket
Monitor recorder
IBP
Label : ART, PA, CVP, RAP, LAP, ICP, P1, P2
Measuring and alarm range
ART: 0-300 mmHg
PA : -6~120mmHg
CVP/RAP/LAP/CP: -10-40 mmHg
P1/P2: -10~300mmHg
Press Sensor
Sensitivity : 5uV/V/mmHg
Impedance: 300-3000Ω
Resolution: 1 mmHg
Accuracy: ±2% or +1mmhg, which great
Actualization interval: about 1 Sec.
Co2
Side/Main stream
Measurement: Infrared radiation absorption technique
Measurement range: 0~99mmhg
Precision : 0~40mmhg ±2mmhg
41~76mmhg ±5mmhg
77~99mmHg ±10mmHg
Resolution: 1mmHg
Pumping rate : 70ml/min, 100ml/min
Accuracy : 15% or 15ml/min big
AwRR measuring range : 0~120BRPM
Precision: 0~70BrPM: ±2BrPM
>70BrPM: ±5BrPM
Note: the specifications are subject to changes without prior notice.