Two Vaporizers Anesthesia Workstation
Product features
Integrated breathing circuit with APL valve, excellent tightness.
·Replacing drive bellow is NOT required when operating Adult/pediatric anesthesia.
·10.1 inch touch screen with no dead space.
·Applied with 2 high-precision vaporizers. (Selectatec).
·Real-time detection for loops, Paw, BPM, VT and compliance.
·Comprehensive monitoring capabilities include ventilation parameters monitoring, preset parameters, alarm indication and breathing waveform monitoring.
·Head light
·Altitude compensation function.
·ACGO available to connect with T/Bain circuit for open operation.
·Yoke system with up to 3 back up cylinders (A type with 11L size)(Optional).
Technical Specification:
Modes of Ventilation | A/C: Assist/control Ventilation VCV: Volume Controlled Ventilation SIMV: Synchronized Intermittent Mandatory Ventilation MANUAL: Manual PCV: Pressure Control Ventilation PSV: Pressure Support Ventilation STAND-BY: STAND-BY |
Ventilator parameter | Working type: Electronically controlled, pneumatic driven Tidal Volume range: 20-1500ml Pressure range: 5-60 cmH2O Respiration Rate Range: 1-100bpm (1-40bpm under SIMV) I: E Range: 4:1-1:8 Trigger Range: -20-20cmH2O(based on PEEP) Flow trigger: 0-15L Minute volume range: >18L/Min SIGH range: 0-5/100 |
System Standard | Hypoxic guard system: N2O cut-off valve, O2 concentration >25% Safety valve: <12.5kPa Fresh gas compensation: 25-75L/Min Flow meter: 5 tubes, cascade 0.1-15L/min O2, 0.1-15L/min N2O, 0.1-15L/min air Volume of CO2 absorber: 2L Vaporizer: 2 Selectable, maximum 2 mounts(Isoflurane/Enflurane/Sevoflurane/Halothane) |
Monitoring | Display graphics: Waveforms of P-T, F-T, V-T, loops of P-V, V-F, F-P Types: VT, MV, BPM, Paw, Lung compliance, Inspiratory Platform, FiO2 |
Alarm Limits | Audio/Visual: No tidal volume, MV, Paw, Fio2 limitation O2, Air failure, AC power failure, Battery low |
PEEP | Type: Integrated, Electronic control Range: OFF, 3-30cmH2O |
Optional | 1.AGSS 2.Suction Unit 3.Communication Interfac |